MassChrom Cortisol, Cortisone Saliva Control Level I, Order No. 0349 - Product Usage: The kit is used as a test for patients to monitor their cortisol and cortisone levels.
Recall
- Recall Number
- Z-0522-2021
- Event Number
- 86699
- Firm
- Chromsystems Instruments & Chemicals GmbH Am Haag 12 Grafelfing Germany
- FEI Number
- 3003951180
- Product Code
- JJY
- Status
- Completed
- Root Cause
- Under Investigation by firm
- Initiated
- October 29, 2020
Description
MassChrom Cortisol, Cortisone Saliva Control Level I, Order No. 0349 - Product Usage: The kit is used as a test for patients to monitor their cortisol and cortisone levels.
Control lots may not meet stability criteria and could result in increased results of patient samples for cortisone.
The firm sent consignees a recall notification letter dated 10/28/2020. Consignees are asked to stop using products and to destroy remaining stock of affected product in accordance with the consignees laboratory procedures. The recalling firm will replace affected stock with new lots. Consignees are asked to further disseminate recall information to any other customers that the product has been shipped to. A response form is included with the recall notification letter and customers are asked to complete the form and return to the recalling firm.
US Nationwide distribution in the states of NC and NY.
880 vials