FDA Recall
Terminated
Ingenuity TF PET/CT - a integrated diagnostic X-ray Computed Tomography (CT) and Positron Emission Tomography (PET) system
Recall: Z-0522-2018
·
Initiated November 1, 2017
Recall
- Recall Number
- Z-0522-2018
- Event Number
- 79004
- Firm
- Philips Medical Systems (Cleveland) Inc
- FEI Number
- 1525965
- Product Code
- KPS
- Status
- Terminated
- Root Cause
- Vendor change control
- Initiated
- November 1, 2017
- Terminated
- August 14, 2018
- Address
- 595 Miner Rd, Cleveland, OH, 44143-2131
Description
Ingenuity TF PET/CT - a integrated diagnostic X-ray Computed Tomography (CT) and Positron Emission Tomography (PET) system
Reason
One of the two wire channels mounted inside the Gantry Separation Unit (GSU) that is also used to secure the top two GSU covers, was missing mounting hardware.
Action
Philips Healthcare is notifying the affected users of this issue via this Customer Information Letter and correcting via Field Change Order FC088200514.
Distribution
One domestic and 1 foreign consignee ( Spain)
Quantity
2