FDA Recall Terminated

Stryker Tom Tessier Osseous Microtome: Plug for Bone Mill Systems (Parts 01-15400 and 01-15401), Stryker Leiblinger Inc., Kalamazoo, MI; Part 01-15407.

Recall: Z-0521-2007 · Initiated December 22, 2006

Recall

Recall Number
Z-0521-2007
Event Number
37067
Firm
Stryker Leibinger USA
FEI Number
3005101424
Product Code
LXH
Status
Terminated
Root Cause
Other
Initiated
December 22, 2006
Posted
February 27, 2007
Terminated
June 4, 2008
Address
750 Trade Center Way, Ste 200, Portage, MI, 49002-0482

Description

Stryker Tom Tessier Osseous Microtome: Plug for Bone Mill Systems (Parts 01-15400 and 01-15401), Stryker Leiblinger Inc., Kalamazoo, MI; Part 01-15407.

Reason

Metal shavings come off of the plug as a result of friction between the plug and the cutting cylinder, and may be mixed with the milled bone chips.

Action

Consignees were notified via recall letter dated 12/22/06 which instructed them to quarantine the product and to contact the firm for further instructions.

Distribution

Nationwide.

Quantity

26