COMP RVRS Shoulder Baseplates, including Custom Products, Item numbers 115331S 115331 115330S 115330 110040610 110040620 010000589 CP561861 PM0001456 PM551047 PM551063 PM555137 PM555160 PM555209 PM555322 PM555406 PM555415 UDI: (01)00880304475342(17)211128(10)704850 (01)00880304475342(17)220128(10)094500 (01)00880304475342(17)220328(10)856620 (01)00880304475342(17)220528(10)487710 (01)00880304475342(17)220628(10)028860 (01)00880304475342(17)220628(10)028870 (01)00880304475342(17)220628(10)799030 (01)00880304475342(17)220628(10)799070 (01)00880304475342(17)220628(10)929900 (01)00880304532465(17)220628(10)850060 (01)00880304475342(17)220628(10)803010 (01)00880304532465(17)220728(10)026780 (01)00880304475342(17)220828(10)208370 (01)00880304475342(17)220901(10)811210R (01)00880304475342(17)221028(10)960160 (01)00880304532465(17)221028(10)943820 (01)00880304475342(17)240616(10)237420 (01)00880304475342(17)230909(10)500790 (01)00880304475342(17)221128(10)469390 (01)00880304475342(17)221128(10)864610 (01)00880304532465(17)221128(10)234970 (01)00880304475342(17)221204(10)038140 (01)00880304475342(17)221212(10)630530 (01)00880304475342(17)221212(10)508750 (01)00880304475342(17)221212(10)769380 (01)00880304475342(17)221213(10)801090 (01)00880304475342(17)221213(10)831150 (01)00880304475342(17)221217(10)812790 (01)00880304475342(17)221218(10)543530 (01)00880304475342(17)221219(10)630010 (01)00880304532465(17)221228(10)458650 (01)00880304532465(17)221228(10)752440 (01)00880304532465(17)221228(10)786080 (01)00880304475342(17)230103(10)913540 (01)00880304475342(17)230103(10)628460 (01)00880304475342(17)230103(10)250450 (01)00880304475342(17)230104(10)913530 (01)00880304475342(17)230104(10)994010 (01)00880304475342(17)230107(10)591420 (01)00880304475342(17)230107(10)913550 (01)00880304475342(17)230108(10)732010 (01)00880304475342(17)230111(10)913570 (01)00880304475342(17)230111(10)035570 (01)00880304475342(17)230114(10)918370 (01)00880304475342(17)230117(10)042060 (01)00880304475342(17)230118(10)042100 (01)00880304475342(17)230118(10)014090 (01)00880304475342(17)230121(10)014110 (01)00880304475342(17)230124(10)081050 (01)00880304475342(17)230128(10)967360 (01)00880304532465(17)230128(10)014050 (01)00880304475342(17)230129(10)967280 (01)00880304475342(17)230204(10)913560 (01)00880304475342(17)230205(10)460190 (01)00880304475342(17)230205(10)207400 (01)00880304475342(17)230206(10)081040 (01)00880304532465(17)230207(10)014060 (01)00880304475342(17)230207(10)081030 (01)00880304532465(17)230207(10)843610 (01)00880304532465(17)230207(10)319470 (01)00880304475342(17)230207(10)081060 (01)00880304475342(17)230207(10)274520 (01)00880304475342(17)230208(10)165560 (01)00880304532465(17)230208(10)165530 (01)00880304475342(17)230211(10)165570 (01)00880304532465(17)230211(10)207450 (01)00880304532465(17)230211(10)520910 (01)00880304532465(17)230214(10)112280 (01)00880304475342(17)230214(10)460470 (01)00880304532465(17)230214(10)555280 (01)00880304532465(17)230214(10)173830 (01)00880304532465(17)230215(10)014040 (01)00880304532465(17)230215(10)387150 (01)00880304475342(17)230215(10)165550 (01)00880304475342(17)230215(10)207430 (01)00880304475342(17)230215(10)319510 (01)00880304532465(17)230215(10)512310 (01)00880304532465(17)230215(10)124710 (01)00880304532465(17)230215(10)274460 (01)00880304532465(17)230215(10)322260 (01)00880304532465(17)230215(10)562180 (01)00880304532465(17)230215(10)006420 (01)00880304532465(17)230218(10)319480 (01)00880304475342(17)230218(10)433460 (01)00880304532465(17)230219(10)433400 (01)00880304475342(17)230219(10)319500 (01)00880304532465(17)230219(10)487040 (01)00880304532465(17)230219(10)491250 (01)00880304475342(17)230220(10)446680 (01)00880304532465(17)230220(10)433430 (01)00880304532465(17)230221(10)446660 (01)00880304532465(17)230222(10)433380 (01)00880304475342(17)230222(10)594640 (01)00880304532465(17)230225(1
Recall
- Recall Number
- Z-0520-2020
- Event Number
- 83783
- Firm
- Zimmer Biomet, Inc.
- FEI Number
- 1825034
- Product Code
- KWS
- Status
- Terminated
- Root Cause
- Environmental control
- Initiated
- August 21, 2019
- Terminated
- July 21, 2021
- Address
- 56 E Bell Dr, Warsaw, IN, 46582-6989
Description
COMP RVRS Shoulder Baseplates, including Custom Products, Item numbers 115331S 115331 115330S 115330 110040610 110040620 010000589 CP561861 PM0001456 PM551047 PM551063 PM555137 PM555160 PM555209 PM555322 PM555406 PM555415 UDI: (01)00880304475342(17)211128(10)704850 (01)00880304475342(17)220128(10)094500 (01)00880304475342(17)220328(10)856620 (01)00880304475342(17)220528(10)487710 (01)00880304475342(17)220628(10)028860 (01)00880304475342(17)220628(10)028870 (01)00880304475342(17)220628(10)799030 (01)00880304475342(17)220628(10)799070 (01)00880304475342(17)220628(10)929900 (01)00880304532465(17)220628(10)850060 (01)00880304475342(17)220628(10)803010 (01)00880304532465(17)220728(10)026780 (01)00880304475342(17)220828(10)208370 (01)00880304475342(17)220901(10)811210R (01)00880304475342(17)221028(10)960160 (01)00880304532465(17)221028(10)943820 (01)00880304475342(17)240616(10)237420 (01)00880304475342(17)230909(10)500790 (01)00880304475342(17)221128(10)469390 (01)00880304475342(17)221128(10)864610 (01)00880304532465(17)221128(10)234970 (01)00880304475342(17)221204(10)038140 (01)00880304475342(17)221212(10)630530 (01)00880304475342(17)221212(10)508750 (01)00880304475342(17)221212(10)769380 (01)00880304475342(17)221213(10)801090 (01)00880304475342(17)221213(10)831150 (01)00880304475342(17)221217(10)812790 (01)00880304475342(17)221218(10)543530 (01)00880304475342(17)221219(10)630010 (01)00880304532465(17)221228(10)458650 (01)00880304532465(17)221228(10)752440 (01)00880304532465(17)221228(10)786080 (01)00880304475342(17)230103(10)913540 (01)00880304475342(17)230103(10)628460 (01)00880304475342(17)230103(10)250450 (01)00880304475342(17)230104(10)913530 (01)00880304475342(17)230104(10)994010 (01)00880304475342(17)230107(10)591420 (01)00880304475342(17)230107(10)913550 (01)00880304475342(17)230108(10)732010 (01)00880304475342(17)230111(10)913570 (01)00880304475342(17)230111(10)035570 (01)00880304475342(17)230114(10)918370 (01)00880304475342(17)230117(10)042060 (01)00880304475342(17)230118(10)042100 (01)00880304475342(17)230118(10)014090 (01)00880304475342(17)230121(10)014110 (01)00880304475342(17)230124(10)081050 (01)00880304475342(17)230128(10)967360 (01)00880304532465(17)230128(10)014050 (01)00880304475342(17)230129(10)967280 (01)00880304475342(17)230204(10)913560 (01)00880304475342(17)230205(10)460190 (01)00880304475342(17)230205(10)207400 (01)00880304475342(17)230206(10)081040 (01)00880304532465(17)230207(10)014060 (01)00880304475342(17)230207(10)081030 (01)00880304532465(17)230207(10)843610 (01)00880304532465(17)230207(10)319470 (01)00880304475342(17)230207(10)081060 (01)00880304475342(17)230207(10)274520 (01)00880304475342(17)230208(10)165560 (01)00880304532465(17)230208(10)165530 (01)00880304475342(17)230211(10)165570 (01)00880304532465(17)230211(10)207450 (01)00880304532465(17)230211(10)520910 (01)00880304532465(17)230214(10)112280 (01)00880304475342(17)230214(10)460470 (01)00880304532465(17)230214(10)555280 (01)00880304532465(17)230214(10)173830 (01)00880304532465(17)230215(10)014040 (01)00880304532465(17)230215(10)387150 (01)00880304475342(17)230215(10)165550 (01)00880304475342(17)230215(10)207430 (01)00880304475342(17)230215(10)319510 (01)00880304532465(17)230215(10)512310 (01)00880304532465(17)230215(10)124710 (01)00880304532465(17)230215(10)274460 (01)00880304532465(17)230215(10)322260 (01)00880304532465(17)230215(10)562180 (01)00880304532465(17)230215(10)006420 (01)00880304532465(17)230218(10)319480 (01)00880304475342(17)230218(10)433460 (01)00880304532465(17)230219(10)433400 (01)00880304475342(17)230219(10)319500 (01)00880304532465(17)230219(10)487040 (01)00880304532465(17)230219(10)491250 (01)00880304475342(17)230220(10)446680 (01)00880304532465(17)230220(10)433430 (01)00880304532465(17)230221(10)446660 (01)00880304532465(17)230222(10)433380 (01)00880304475342(17)230222(10)594640 (01)00880304532465(17)230225(1
Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue.
On September 11, 2019, the firm began notifying distributors and customers of the recall via an Urgent Medical Device Recall letter. The letter informed consignees of the product issue. Customers were asked to assist their Zimmer Biomet sales representative and quarantine all identified product. The sales representative will remove the product from your facility. If any products were further distributed, please forward the notification to responsible parties. If you have further questions or concerns, please call customer service at 574-371-3071 between 8:00 am and 5:00 pm ET, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected]. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule.
US Nationwide distribution and countries of Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, El Salvador, Italy, Netherlands, India, Japan, Korea, Malaysia, Mexico, New Zealand, Panama, Singapore, Taiwan, Thailand, Trinidad and Tobago, and Venezuela.
219988 units