Cook brand Zenith Flex AAA Endovascular graft with the H&L-B One-Shot Introduction System( A= 30mm, B=155mm, C=125mm, D=12mm, E= 20.0f, F=.035'', G =40cm); Reorder number TFBB-30-125.
Recall: Z-0520-05
·
Initiated December 20, 2004
Recall
- Recall Number
- Z-0520-05
- Event Number
- 30721
- Firm
- Cook, Inc.
- FEI Number
- 1820334
- Product Code
- MIH
- Status
- Terminated
- Root Cause
- Other
- Initiated
- December 20, 2004
- Posted
- February 3, 2005
- Terminated
- December 2, 2005
- Address
- 750 Daniels Way, Bloomington, IN, 47404-9120
Description
Cook brand Zenith Flex AAA Endovascular graft with the H&L-B One-Shot Introduction System( A= 30mm, B=155mm, C=125mm, D=12mm, E= 20.0f, F=.035'', G =40cm); Reorder number TFBB-30-125.
Reason
The firm has received complaints of difficulty in releasing the barbed suprarenal stent from the top cap that constrains it, which may result in misplacement of the graft.
Action
Recall letter dated 12/20/04 was issued to each consignee via FedEx instructing them to cease using the products and to return them to the firm.
Distribution
Nationwide and Canada.