FDA Recall Open, Classified

Para-Pak Clean Vial-For the collection, transportation, preservation, and examination of stool specimens containing intestinal parasites Catalog Number: 900312

Recall: Z-0519-2024 · Initiated October 19, 2023

Recall

Recall Number
Z-0519-2024
Event Number
93350
Firm
Meridian Bioscience Inc
FEI Number
1524213
Product Code
FMH
Status
Open, Classified
Root Cause
Process control
Initiated
October 19, 2023
Posted
December 8, 2023
Address
3471 River Hills Dr, Cincinnati, OH, 45244-3023

Description

Para-Pak Clean Vial-For the collection, transportation, preservation, and examination of stool specimens containing intestinal parasites Catalog Number: 900312

Reason

Plastic vials contain a manufacturing defect may cause an increased risk of leakage and presents a risk of incidental exposure to biological hazards associated with direct contact with the patient specimen.

Action

Meridan issued Urgent Medical Device Correction letters via email on 10/19/23 to Distributors and End-Users. Letter states reason for recall, health risk and action to take: Identify and segregate any remaining inventory. REQUIRED ACTIONS FOR PARA-PAK CLEAN VIAL, CATALOG 900312, LOT 509940Q: " Identify and segregate any remaining inventory. " Discard remaining Inventory of Para-Pak Clean Vial, Catalog 900312, Lot 509940Q in accordance with local, state, and/or federal regulations. Record quantity discarded. o Contact Meridian Technical Support at 1-800-343-3858 to coordinate product replacement activities. " Complete and return the Customer Notification Form. This will indicate notification of this field correction. Complete this form even if you have no remaining inventory. REQUIRED ACTIONS FOR PARA-PAK Zn-PVA/10% FORMALIN, CATALOG 301012, LOT 509130Q: " Identify and segregate any remaining inventory. " For remaining inventory of Para-Pak Zn-PVA/10% Formalin, Catalog 301012, Lot 509130Q, inspect each individual vial set for signs of leakage. o Leakage would be able to be visually observed through inspection of the Ziplock-style bag containing the two vials of preservatives. Signs of leakage include identification of preservative outside of the plastic vials or at the bottom of the Ziplock-style bag. DO NOT remove vials that exhibit leaking from the Ziplock-style bag. Handle with care to prevent direct contact of the preservative with the skin or eyes. o Record the quantity of 2 vial sets that demonstrated signs of leakage. " Discard the remaining inventory of Para-Pak Zn-PVA/10% Formalin, Catalog 301012, Lot 509130Q in accordance with local, state, and/or federal regulations. Record quantity discarded. o Contact Meridian Technical Support 1-800-343-3858 to coordinate product replacement activities. " Complete and return the Customer Notification Form. This will indicate notification of this field correction. Complete this form even if you have no remaining invento

Distribution

Nationwide Foreign: Italy

Quantity

571 Cases