The GISH Vision Hollow Fiber Oxygenator with GBS coating (a non-leaching heparin coating) is a sterile, non-pyrogenic, single use, disposable, extracorporeal device designed to be used in an extracorporeal blood circuit to control exchange of gases. It is intended for the supply of oxygen to, and the removal of carbon dioxide from, the blood of humans. An integral heat exchanger is included which is intended to control the temperature of the blood.
Recall
- Recall Number
- Z-0519-06
- Event Number
- 34215
- Firm
- Gish Biomedical Inc
- FEI Number
- 2021836
- Product Code
- DTZ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- November 28, 2005
- Posted
- February 14, 2006
- Terminated
- June 21, 2007
- Address
- 22942 Arroyo Vis, Rancho Santa Margarita, CA, 92688-2600
Description
The GISH Vision Hollow Fiber Oxygenator with GBS coating (a non-leaching heparin coating) is a sterile, non-pyrogenic, single use, disposable, extracorporeal device designed to be used in an extracorporeal blood circuit to control exchange of gases. It is intended for the supply of oxygen to, and the removal of carbon dioxide from, the blood of humans. An integral heat exchanger is included which is intended to control the temperature of the blood.
Gish received four reports of potential leakage. All reports were associated with GBS coating. There was no patient injury reported.
The firm faxed all initial consignees a letter notifying them of the recall and the affected lots involved. 11/14/2005 The consignee was instructed to immediately inspect their inventory, and if any units of the affected product remain, the consignee is to contact Gish Biomedical Customer Service at 866-221-9911. The consignee is to provide catalogue number, lot number and quantity and state the reason for replacement is the recall letter. Customer Service will, at the same time, issue a SORMA for the return of coated oxygenators from the consignee. Sequester the coated oxygenator for return to Gish Biomedical. If the consignees prefer, they can provide a contact name and the number of packages to be picked up and the firm will issue a Federal Express call tag for pick-up at the earliest scheduled time.
Nationwide (CA, FL, IL, IN, MA, MI, MS, NH, NJ, NY, OH, PA, WI, VA), Canada, and Mexico.
2,306