FDA Recall
Terminated
Magnetom Espree with Swiveling OR (Operating Room) Table, Magnetic Resonance Imaging System,Siemens Medical Solutions USA, Inc., Malvern, PA
Recall: Z-0518-2008
·
Initiated November 5, 2007
Recall
- Recall Number
- Z-0518-2008
- Event Number
- 45986
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- LNH
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- November 5, 2007
- Posted
- February 5, 2008
- Terminated
- February 5, 2008
- Address
- 51 Valley Stream Pkwy, Malvern, PA, 19355-1406
Description
Magnetom Espree with Swiveling OR (Operating Room) Table, Magnetic Resonance Imaging System,Siemens Medical Solutions USA, Inc., Malvern, PA
Reason
Table may experience a deadlock situation
Action
The recalling firm issued a Customer Safety Advisory to affected customers per Update Instructions MR031/07/S. The letter informs customers of the potential issue and provides instructions to avoid its occurrence. The recalling firm has also released the software update to correct the issue via Update Instructions MR034/07/S. Notifications were issued on or about November 5, 2007.
Distribution
The products were shipped to medical facilities in TX and NY.
Quantity
4 units