FDA Recall
Terminated
Quickie brand Freestyle M11 & F11, C. G. Tilt, Powered Wheelchair; Product is manufactured and distributed by Sunrise Medical, 2842 Business Park Ave., Fresno, CA 93727
Recall: Z-0518-06
·
Initiated November 15, 2005
Recall
- Recall Number
- Z-0518-06
- Event Number
- 34155
- Firm
- Sunrise Medical Inc
- FEI Number
- 2937137
- Product Code
- ITI
- Status
- Terminated
- Root Cause
- Other
- Initiated
- November 15, 2005
- Posted
- February 14, 2006
- Terminated
- December 28, 2006
- Address
- 2842 N Business Park Ave, Fresno, CA, 93727-1328
Description
Quickie brand Freestyle M11 & F11, C. G. Tilt, Powered Wheelchair; Product is manufactured and distributed by Sunrise Medical, 2842 Business Park Ave., Fresno, CA 93727
Reason
While operating the powered wheelchair beyond a specified angle of recline, the wheelchair can travel at an unsafe speed when in this position.
Action
On November 9, 2005, the firm contacted its consignees via telephone, which was followed up by a mailing, to alert them of the situation. The firm is making arrangements to conduct the service correction at users'' residence. All appointmens are scheduled to be completed by December 15, 2005.
Distribution
All 5 units were distributed to direct consignees, in CA, MA, NC, and OH.
Quantity
5 units