FDA Recall Terminated

OEC Uroview 2800 with Dual Monitor Option, fluoroscopic x-ray system, Model Numbers: 882082-01, 882082-02, 886765-01, GE Healthcare, Surgery, Salt Lake City, UT.

Recall: Z-0514-2007 · Initiated January 15, 2007

Recall

Recall Number
Z-0514-2007
Event Number
37191
Firm
GE OEC Medical Systems, Inc
FEI Number
1720753
Product Code
JAA
Status
Terminated
Root Cause
Other
Initiated
January 15, 2007
Posted
February 15, 2007
Terminated
December 9, 2011
Address
384 Wright Brothers Dr, Salt Lake City, UT, 84116-2862

Description

OEC Uroview 2800 with Dual Monitor Option, fluoroscopic x-ray system, Model Numbers: 882082-01, 882082-02, 886765-01, GE Healthcare, Surgery, Salt Lake City, UT.

Reason

Uroview 2800 system with dual monitors may pose a safety risk to patients if the monitors become disconnected and fall.

Action

Consignees were notified by letter on 01/15/2007. Instructions were given to position and not move the dual monitors until inspection and correction by GE personnel. For the affected units with a 3 phase power distribution box, users were cautioned to ensure that the access door to the power box is completely closed when the unit is powered on to prevent a shock hazard.

Distribution

Worldwide, including USA, Canada, China, France, Germany, Hong Kong, Italy, Japan, Singapore, South Africa, Switzerland, and Venezuela.

Quantity

425 units