FDA Recall Terminated

PerkinElmer QSight 210 MD Mass Spectrometer Simplicity 3Q MD v1.0 with Hot Fix 2 Software intended to identify compounds in human samples by ionizing the compound and separating the resulting ions. Part Number: BC004872

Recall: Z-0513-2019 · Initiated September 24, 2018

Recall

Recall Number
Z-0513-2019
Event Number
81266
Firm
PerkinElmer Health Sciences Canada, Inc.
FEI Number
3013944243
Product Code
DOP
Status
Terminated
Root Cause
Software design
Initiated
September 24, 2018
Terminated
September 17, 2019
Address
32 Nixon Rd, Unit 1 Bolton Canada

Description

PerkinElmer QSight 210 MD Mass Spectrometer Simplicity 3Q MD v1.0 with Hot Fix 2 Software intended to identify compounds in human samples by ionizing the compound and separating the resulting ions. Part Number: BC004872

Reason

Under certain conditions when using PerkinElmer QSight 210 MD mass spectrometer with Simplicity 3Q MD and Hotfix 2 software, it is possible for the original data from the last acquired sample to be over-written when the automatic flush is enabled.

Action

Perkin Elmer issued notification on 9/24/18 customers are requested to follow instructions contained within the Urgent Medical Device Correction letter and disseminate the information to appropriate functions within their organization. Inform relevant personnel and execute one of the following two options: Qption 1 (Users who do not require Autamatic Flushinq): Deactivate the automatic flushing option in your software: 1) Navigate to the Batch Acquisition Queue tab 2) Click the Queue Options button 3) Under the "System Flush and Standby" section, DESELECT the "Enable automatic flushing" option. Option 2 (Users who require Automatic Flushinqt: Add a non data (blank or "dummy") sample to the acquisition queue to be run as your final sample (before going to Flush/standby). This will ensure that a true sample or control data are not overwritten. Contact PerkinElmer Heatth Scbnces C,arda, Bolton facility Quality Manager- SandiOrdog for further information (647. 505. 8682). The issue identified above will be addressed through a software update. PerkinElmer will contact you when the fix is available and upgrade your software.

Distribution

FL, MN, PA India, Italy, Korea, Spain

Quantity

18 units