FDA Recall Terminated

Siemens Healthcare Diagnostics, Dimension Vista(R) Cardiac Troponin I Calibrator (KC678) lot 1DD085, in vitro diagnostic

Recall: Z-0513-2012 · Initiated November 17, 2011

Recall

Recall Number
Z-0513-2012
Event Number
60643
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2517506
Product Code
JIT
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
November 17, 2011
Posted
January 11, 2012
Terminated
August 26, 2014
Address
500 GBC Drive PO BOX 6101, Mailstop 514, Newark, DE, 19714-6101

Description

Siemens Healthcare Diagnostics, Dimension Vista(R) Cardiac Troponin I Calibrator (KC678) lot 1DD085, in vitro diagnostic

Reason

Siemens Healthcare Diagnostics has confirmed positive bias on QC and Patients.

Action

Siemens healthcare Diagnostics initiated their recall of lot 1DD085 on November 17, 2011 by sending an Urgent Field Safety Notice letter to all affected customers. Customers were requested to immediately discontinue use and discard any remaining inventory of the CTNI CAL lot 1DD085.

Distribution

Worldwide distribution, including USA, Austria, Belgium, Canada, Denmark, Finland, France, Germany, Italy, Japan, Netherlands, New Zealand, Portugal, Puerto Rico, Slovenia, South Korea, Spain and Switzerland.

Quantity

2281 cartons