FDA Recall
Terminated
Siemens Healthcare Diagnostics, Dimension Vista(R) Cardiac Troponin I Calibrator (KC678) lot 1DD085, in vitro diagnostic
Recall: Z-0513-2012
·
Initiated November 17, 2011
Recall
- Recall Number
- Z-0513-2012
- Event Number
- 60643
- Firm
- Siemens Healthcare Diagnostics, Inc.
- FEI Number
- 2517506
- Product Code
- JIT
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- November 17, 2011
- Posted
- January 11, 2012
- Terminated
- August 26, 2014
- Address
- 500 GBC Drive PO BOX 6101, Mailstop 514, Newark, DE, 19714-6101
Description
Siemens Healthcare Diagnostics, Dimension Vista(R) Cardiac Troponin I Calibrator (KC678) lot 1DD085, in vitro diagnostic
Reason
Siemens Healthcare Diagnostics has confirmed positive bias on QC and Patients.
Action
Siemens healthcare Diagnostics initiated their recall of lot 1DD085 on November 17, 2011 by sending an Urgent Field Safety Notice letter to all affected customers. Customers were requested to immediately discontinue use and discard any remaining inventory of the CTNI CAL lot 1DD085.
Distribution
Worldwide distribution, including USA, Austria, Belgium, Canada, Denmark, Finland, France, Germany, Italy, Japan, Netherlands, New Zealand, Portugal, Puerto Rico, Slovenia, South Korea, Spain and Switzerland.
Quantity
2281 cartons