TomoTherapy Treatment System with software versions 2.0.1 / 2.0.2 / 2.0.3/ (Hi:Art 5.0.1 / 5.0.2 / 5.0.3)
Recall
- Recall Number
- Z-0507-2015
- Event Number
- 69854
- Firm
- TomoTherapy Incorporated
- FEI Number
- 3003873069
- Product Code
- IYE
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- November 21, 2014
- Posted
- December 12, 2014
- Terminated
- July 17, 2015
- Address
- 1240 Deming Way, Madison, WI, 53717-1954
Description
TomoTherapy Treatment System with software versions 2.0.1 / 2.0.2 / 2.0.3/ (Hi:Art 5.0.1 / 5.0.2 / 5.0.3)
Accuray has become aware of a potential safety issue related to the TomoTherapy Treatment System caused by a failure to monitor the jaw position after a jaw error occurs. This may result in an incorrect jaw position during treatment, without generating system interruption.
Accuray sent an "Urgent Field Safety Notice" dated November 21, 2014, to all affected customers. The letter identified the problem the product and the action needed to be taken by the customers. For questions they can contact Accuray Customer Support at 1-877-668-8667 or visit the Accuray Technical Solutions Center at www.accuray.com>Services & Support or call: North America: +1.877.668.8667, Europe: +800.4141.9595, Hong Kong: +852.2247.8688, Japan: +81.3.6269.9556. Requested consignees to complete and return the Acknowledgement Form to Accuray Regulatory Affairs.
Worldwide Distribution - USA (nationwide) included the states of : AL, AZ, AR, CA, FL, GA, IL, IN, KS, ,Y, MD, MN, MO, NE, NM, NY, NC, OK, OR, PA, SD, TN, TX, UT, WA, WV, WI and WY., and the countries of : JAPAN, BELGIUM, NETHERLANDS, ITALY, GERMANY, KOREA, FRANCE, SWITZERLAND, ENGLAND, CANADA, MEXICO, HONG KONG, AUSTRALIA, SRI LANKA, GREECE, CHINA, PHILIPPINES, MARTINIQUE, ROMANIA, SPAIN and TURKEY.
210 ( 72 USA; 138 OUS )