FDA Recall Terminated

TomoTherapy Treatment System with software versions 2.0.1 / 2.0.2 / 2.0.3/ (Hi:Art 5.0.1 / 5.0.2 / 5.0.3)

Recall: Z-0507-2015 · Initiated November 21, 2014

Recall

Recall Number
Z-0507-2015
Event Number
69854
Firm
TomoTherapy Incorporated
FEI Number
3003873069
Product Code
IYE
Status
Terminated
Root Cause
Software design
Initiated
November 21, 2014
Posted
December 12, 2014
Terminated
July 17, 2015
Address
1240 Deming Way, Madison, WI, 53717-1954

Description

TomoTherapy Treatment System with software versions 2.0.1 / 2.0.2 / 2.0.3/ (Hi:Art 5.0.1 / 5.0.2 / 5.0.3)

Reason

Accuray has become aware of a potential safety issue related to the TomoTherapy Treatment System caused by a failure to monitor the jaw position after a jaw error occurs. This may result in an incorrect jaw position during treatment, without generating system interruption.

Action

Accuray sent an "Urgent Field Safety Notice" dated November 21, 2014, to all affected customers. The letter identified the problem the product and the action needed to be taken by the customers. For questions they can contact Accuray Customer Support at 1-877-668-8667 or visit the Accuray Technical Solutions Center at www.accuray.com>Services & Support or call: North America: +1.877.668.8667, Europe: +800.4141.9595, Hong Kong: +852.2247.8688, Japan: +81.3.6269.9556. Requested consignees to complete and return the Acknowledgement Form to Accuray Regulatory Affairs.

Distribution

Worldwide Distribution - USA (nationwide) included the states of : AL, AZ, AR, CA, FL, GA, IL, IN, KS, ,Y, MD, MN, MO, NE, NM, NY, NC, OK, OR, PA, SD, TN, TX, UT, WA, WV, WI and WY., and the countries of : JAPAN, BELGIUM, NETHERLANDS, ITALY, GERMANY, KOREA, FRANCE, SWITZERLAND, ENGLAND, CANADA, MEXICO, HONG KONG, AUSTRALIA, SRI LANKA, GREECE, CHINA, PHILIPPINES, MARTINIQUE, ROMANIA, SPAIN and TURKEY.

Quantity

210 ( 72 USA; 138 OUS )