FDA Recall Terminated

WANG Transbronchial Aspiration Needles Cat. No. MW-221, WANG Cytology Needles, Central or Peripheral Regions, 21 g x 15 mm long needle. --- Common Name: Bronchoscope (flexible or rigid) and accessories. The ConMed WANG Transbronchial Aspiration Needles are designed for use with a bronchoscope. They are used to puncture the tracheobronchial wall and to aspirate sufficient tissue and/or cell specimens to stage bronchogenic carcinoma. These are disposable sterile, single-use devices.

Recall: Z-0505-2013 · Initiated November 20, 2012

Recall

Recall Number
Z-0505-2013
Event Number
63728
Firm
ConMed Corporation
FEI Number
3001451571
Product Code
EOQ
Status
Terminated
Root Cause
Component design/selection
Initiated
November 20, 2012
Posted
December 10, 2012
Terminated
August 25, 2015
Address
525 French Rd, Utica, NY, 13502-5945

Description

WANG Transbronchial Aspiration Needles Cat. No. MW-221, WANG Cytology Needles, Central or Peripheral Regions, 21 g x 15 mm long needle. --- Common Name: Bronchoscope (flexible or rigid) and accessories. The ConMed WANG Transbronchial Aspiration Needles are designed for use with a bronchoscope. They are used to puncture the tracheobronchial wall and to aspirate sufficient tissue and/or cell specimens to stage bronchogenic carcinoma. These are disposable sterile, single-use devices.

Reason

ConMed received complaints from a user facility reporting that the WANG Transbronchial Aspiration Needles would not insert properly through the instrument channel of a 2.0 mm bronchoscope. If the device will not fit through the instrument channel of a 2.0 mm bronchoscope, it must be replaced with an alternate device and could result in minor inconvenience to the user.

Action

ConMed Corporation sent an Urgent Field Safety Notice letter dated November 30, 2012 with Response Form , to all affected customers via USPS Priority Mail. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to complete Attachment II and return it with any unused affected product to: ConMed Corporation 525 French Road, Utica, NY 13502 United Sates of America For information regarding this recall call 315-624-3225.

Distribution

Worldwide Distribution - USA including CA , FL, IA, IN, MA, MD, ME, MO, MS, NC, NM, OH, PA, TX. Foreign distribution to Spain, Portugal, Japan and Hungary.

Quantity

Total 452