FDA Recall Terminated

Sinu-Lift(TM) System, LLC., Model SLS-1, Sterile, Innovative Implant Technology, The Sinu-Lift(TM) System is intended to : - Facilitate access to the maxillary sinus, - Lift the sinus membrane from the alveolar ridge of the maxilla with a minimum height of 5mm, and - Augment the space created by lifting the maxillary sinus membrane with bone graft to supplement the maxillary alveolar crest with additional height needed for anchoring an implant.

Recall: Z-0505-2009 · Initiated September 30, 2008

Recall

Recall Number
Z-0505-2009
Event Number
50168
Firm
Innovative Implant Technology
FEI Number
3006005748
Product Code
NDP
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
September 30, 2008
Posted
January 9, 2009
Terminated
October 1, 2009
Address
1825 Main St, Weston, FL, 33326-3683

Description

Sinu-Lift(TM) System, LLC., Model SLS-1, Sterile, Innovative Implant Technology, The Sinu-Lift(TM) System is intended to : - Facilitate access to the maxillary sinus, - Lift the sinus membrane from the alveolar ridge of the maxilla with a minimum height of 5mm, and - Augment the space created by lifting the maxillary sinus membrane with bone graft to supplement the maxillary alveolar crest with additional height needed for anchoring an implant.

Reason

Possible device failure: The cutter's shaft of the Sinu-Drill of the Sinu-Lift System could potentially get stuck inside the tap, thereby not allowing the cutter to spring out of the tap upon encountering the sinus membrane, increasing the possibility of perforating the sinus membrane.

Action

All consignees will be notified by letter entitled "URGENT: MEDICAL DEVICE RECALL" dated September 30, 2008, using priority mail. Consignees will also be contacted by telephone. Consignees will be asked to return the product using the prepaid FedEx label included with the letter. The returned product will be quarantined. For further information, please telephone Innovative Implant Technology at 305-975-4101

Distribution

Worldwide Distribution --- including USA and countries of Canada, Mexico, Japan, Australia, Japan, Egypt, and Turkey.

Quantity

299