FDA Recall
Terminated
Stille Sonesta Video Fluoroscopy Table 6210; manufactured by Stille AB, Solna, Sweden.
Recall: Z-0504-2007
·
Initiated November 9, 2006
Recall
- Recall Number
- Z-0504-2007
- Event Number
- 36855
- Firm
- Stille-Sonesta, Inc
- FEI Number
- 1000516389
- Product Code
- IXR
- Status
- Terminated
- Root Cause
- Other
- Initiated
- November 9, 2006
- Posted
- February 15, 2007
- Terminated
- June 20, 2007
- Address
- 2220 Canton St, Apt 209, Dallas, TX, 75201-5932
Description
Stille Sonesta Video Fluoroscopy Table 6210; manufactured by Stille AB, Solna, Sweden.
Reason
Bolts may back out of their joints causing the back section of the table to tilt.
Action
Notifications letters sent to consignees on 11/20/06. Users are instructed to check the bolts located under the table. The Notice includes instructions with pictures for checking the bolts and tightening them if necessary. Stille AB has developed a kit for replacing the bolt that is involved and will send a service engineer out to the site to do the replacement.
Distribution
Nationwide.
Quantity
66 units.