FDA Recall Open, Classified

Cooper Surgical H/S ELLIPTOSPHERE CATH 5 FR Set-For administering contrast media during Hysterosalpingography or Hysterosonography procedures to detect uterine pathology such as polyps, fibroids, adhesions or endometrial thickening, and/or patency of fallopian tubes Model: 61-4005

Recall: Z-0503-2024 · Initiated March 31, 2022

Recall

Recall Number
Z-0503-2024
Event Number
93239
Firm
CooperSurgical, Inc.
FEI Number
1216677
Product Code
LKF
Status
Open, Classified
Root Cause
Component design/selection
Initiated
March 31, 2022
Posted
December 6, 2023
Address
95 Corporate Dr, Trumbull, CT, 06611-1350

Description

Cooper Surgical H/S ELLIPTOSPHERE CATH 5 FR Set-For administering contrast media during Hysterosalpingography or Hysterosonography procedures to detect uterine pathology such as polyps, fibroids, adhesions or endometrial thickening, and/or patency of fallopian tubes Model: 61-4005

Reason

Incorrect bonding material used for assembly of the catheter, may fail, resulting in detachment of component parts, broken components, or leakage. This breakage may result in delay of procedure

Action

CooperSurgical issued Urgent Medical Device Safety Notice dated March 22, 2023 via UPS Certified Delivery Mail. Letter states reason for recall, health risk and action to take:Our records indicate that you may have purchased the affected Product from CooperSurgical. Please inspect stock and complete the attached Acknowledgement and Receipt Form. Once completed please return the form to CooperSurgical to acknowledge receipt of the notice. If you do not have affected Product in inventory, please use the same enclosed Form to indicate that and return it to CooperSurgical so that we may document receipt of this Recall Notice. A corrective action has been initiated to ensure this failure does not reoccur. We sincerely apologize for any inconvenience caused by this notice. CooperSurgical is committed to high quality, safe and effective products. Please feel free to reach us at 203-601-5200 ext. 3300 or via email at [email protected].

Distribution

Nationwide Foreign: BELGIUM CANADA MALAYSIA NETHERLANDS SPAIN SWEDEN SWITZERLAND

Quantity

169 units