Maxi Move, Model NDA0200-20 Batteries Pack; BHM Medical Inc., 2001 Tanguay, Magog, QC, Canada J1X 5Y5 The equipment is intended to be used in hospitals, nursing homes, or other health care facilities by trained caregivers for the transfer of patients.

Recall: Z-0503-2009 · Initiated November 5, 2008

Recall

Recall Number
Z-0503-2009
Event Number
50151
Firm
Arjo, Inc.
FEI Number
1419652
Product Code
FSA
Status
Terminated
Root Cause
Component design/selection
Initiated
November 5, 2008
Posted
December 16, 2008
Terminated
January 5, 2010
Address
50 Gary Ave, Roselle, IL, 60172-1605

Description

Maxi Move, Model NDA0200-20 Batteries Pack; BHM Medical Inc., 2001 Tanguay, Magog, QC, Canada J1X 5Y5 The equipment is intended to be used in hospitals, nursing homes, or other health care facilities by trained caregivers for the transfer of patients.

Reason

The battery pack NDA0200-20, lot 070826D, sold with the Maxi Move patient lift or as replacement parts, has a defective connector that could lead to the inability of the battery pack to recharge and/or short circuit, resulting in smoke emission.

Action

Arjo sent the Urgent Device Recall Customer Notification dated 11/5/08 and Field Safety Notice (FSN) dated 10/31/08 to all affected Maxi Move customers on 11/5/08. The accounts were informed of the problem with the connector of the battery pack NDA0200-20, lot 070826D, sold with the Maxi Move patient lift (model numbers starting with KMC)or as replacement parts. They instructed the accounts to discontinue use of the battery pack if it does not recharge or has an intermittent electrical contact. The accounts were requested to complete and return to Arjo Inc., the enclosed customer response form indicating that they have received the notification package. The accounts will be contacted by an Arjo representative within a week to arrange the replacement of the battery pack and corresponding connectors (male part/plug) on the lift and battery charger.

Distribution

Nationwide Distribution

Quantity

273 battery packs