Alphatec Spine Solanas Titanium Pedicle Screw, Part Number 63035-12
Recall: Z-0502-2010
·
Initiated August 5, 2009
Recall
- Recall Number
- Z-0502-2010
- Event Number
- 53144
- Firm
- Alphatec Spine, Inc.
- FEI Number
- 2027467
- Product Code
- KWP
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- August 5, 2009
- Posted
- December 14, 2009
- Terminated
- June 7, 2010
- Address
- 5818 El Camino Real, Carlsbad, CA, 92008-8816
Description
Alphatec Spine Solanas Titanium Pedicle Screw, Part Number 63035-12
Reason
Alphatec Spine discovered that the affected lots were not manufactured correctly in that the screw body portion of the assembly has only 22.5 degrees of angulation in the east-west plane instead of the designed 38-40 degrees.
Action
Alphatec Spine telephoned consignees informing them of the above mentioned Public Reason for Recall on August 5, 2009. They were asked to check their inventory to see where this product is located. If the product had not already been used Alphatec arranged to have Federal Express contact consignee for pick-up of the product(s). They were told to expect to receive the replacement(s) within 2-3 days.
Distribution
Nationwide distribution.
Quantity
163 Units