FDA Recall Terminated

STA Neoplastine CI Plus 10; Product Catalogue Number: 0667 Manufactured by Diagnostica Stago, Asnieres, France.

Recall: Z-0502-2009 · Initiated October 20, 2008

Recall

Recall Number
Z-0502-2009
Event Number
50125
Firm
Diagnostica Stago, Inc.
FEI Number
2245451
Product Code
GJS
Status
Terminated
Root Cause
Process control
Initiated
October 20, 2008
Posted
December 18, 2008
Terminated
June 24, 2010
Address
5 Century Dr, Parsippany, NJ, 07054-4607

Description

STA Neoplastine CI Plus 10; Product Catalogue Number: 0667 Manufactured by Diagnostica Stago, Asnieres, France.

Reason

Internal investigation has determined the potential lack of homogeneity between product vials.

Action

Product Recall Notice letters were sent to all customers on October 20, 2008 vis US mail. The letter stated the reason for the notice, the corrective action, and replacement process. Customers are to confirm that they received the Recall Notice by completing the attached "Product Recall Response Form" and return by fax to the Regulatory Affairs Department at 973-867-6028 or 973-695-0095. The firm is prepared to immediately implement the following corrective action: 1) perform QC testing in accordance to directions from the package insert on each vial and destroy vial and replace inventory if determined product was unsatisfactory OR 2) return inventory for replacement with new lots of inventory. If customers have any questions or need any technical assistance, they are to call the DSI Hotline at 1-800-222-2624, option 1.

Distribution

Nationwide Distribution.

Quantity

31,207 units