FDA Recall Open, Classified

OrganOx metra -Intended as a Normothermic Machine Perfusion system of donor transport livers. Model/Catalog Number: D0003 The OrganOx metra is a transportable device intended to be used to sustain donor livers destined for transplantation in a functioning state for a total preservation time of up to 12 hours.

Recall: Z-0501-2025 · Initiated October 15, 2024

Recall

Recall Number
Z-0501-2025
Event Number
95581
FEI Number
3011560054
Product Code
QQK
Status
Open, Classified
Root Cause
Device Design
Initiated
October 15, 2024
Posted
November 21, 2024
Address
OrganOx Ltd Magdalen Centre 1 Robert Robinson Avenue Oxford United Kingdom

Description

OrganOx metra -Intended as a Normothermic Machine Perfusion system of donor transport livers. Model/Catalog Number: D0003 The OrganOx metra is a transportable device intended to be used to sustain donor livers destined for transplantation in a functioning state for a total preservation time of up to 12 hours.

Reason

Battery movement led to damage of one of the battery cells causing it to short to the battery retention bracket

Action

OrganOx issued Important Field Correction Notice letter on 10/15/24 via Fed'X. letter states reason for recall, health risk and action to take: As a result of the investigation into this event, OrganOx has implemented a secondary battery retention mechanism to prevent battery movement due to jarring or impact that could be related to machine transport or movement. The secondary battery retention mechanism will be installed on your machine by OrganOx service personnel no later than 15 December 2024. Recommendations: The metra in your possession continues to be safe and effective in anticipation of this correction and the machine may be used as normal. Should an unexpected rapid decrease in battery voltage occur as noted by alarm from the metra, please determine the impact to the perfusion event and take action if needed. Should this occur, please follow the OrganOx complaints reporting process and quarantine the machine until it can be serviced. Should you have questions about this important Field Correction please contact: Michael Donahue [email protected] +1 732-948-2175

Distribution

California, Texas, Ohio, Florida, New York, Massachusetts, Virginia, Oklahoma, DC, South Carolina and Georgia.

Quantity

20 units