FDA Recall Terminated

Zimmer Reconstruction System Offset Clamp Tips 4.5 cm Offset, Item # 00-1179-028-03, Zimmer Inc., Warsaw, IN. Intended to temporarily reduce pelvic fractures for fixation during reconstructive surgery.

Recall: Z-0501-2011 · Initiated February 16, 2010

Recall

Recall Number
Z-0501-2011
Event Number
55119
Firm
Zimmer Inc.
FEI Number
1000220733
Product Code
HWN
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
February 16, 2010
Posted
December 1, 2010
Terminated
October 5, 2011
Address
345 E Main St, Warsaw, IN, 46580-2746

Description

Zimmer Reconstruction System Offset Clamp Tips 4.5 cm Offset, Item # 00-1179-028-03, Zimmer Inc., Warsaw, IN. Intended to temporarily reduce pelvic fractures for fixation during reconstructive surgery.

Reason

The firm determined that the recalled instruments were manufactured using the wrong grade of material. As manufactured, the instruments are more brittle, increasing the potential for fracture.

Action

Zimmer sent an Urgent Device Recall letter dated February 16, 2010, to all affected customers. The letter identified the product, the problem, and the action the customers should take. Customers were instructed to monitor all potentially affected patients through regular office consultations. Ensure Zimmer Reconstruction System Tenaculum System Instrument Users at their facility were informed of the urgent device recall. Locate all instruments and quarantine them immediately. Count all affected product in their territory and record the data on the Inventory Return Certification Form included with the letter. Fax a copy of the completed Inventory Return Form to: Zimmer, Inc. at (574) 372-4265. Return the recalled product along with the original completed Inventory Return Certification Form to: Zimmer Distribution Center Attn: Product Service 1777 West Center Street Warsaw, IN 46580 For questions customers were instructed to contact their Zimmer Sales Representative. For questions regarding this recall call 1 - (800) 613-6131.

Distribution

Worldwide Distribution - USA including AL, FL, HI, IA, IL, KS KY, LA, MI, MO, MS, NC, ND, NJ, OH, OK, OR, PA, TN, TX, UT, VA, WA, AND WI and the countries of Australia, Canada, Central or South America, China, Germany, India, Italy, Saudi Arabia, and UK,

Quantity

217 all products