FDA Recall Terminated

Dimension Vista Lactate Dehydrogenase (LDI) Flex reagent cartridge, an in vitro diagnostic test for the quantitative measurement of lactate dehydrogenase in human serum and plasma on the Dimension Vista System.Cat. No. K2054, Mat. No. 10464323

Recall: Z-0499-2019 · Initiated October 11, 2018

Recall

Recall Number
Z-0499-2019
Event Number
81433
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2517506
Product Code
CFJ
Status
Terminated
Root Cause
Process control
Initiated
October 11, 2018
Terminated
June 29, 2020
Address
500 Gbc Dr Ms 514, Newark, DE, 19702-2466

Description

Dimension Vista Lactate Dehydrogenase (LDI) Flex reagent cartridge, an in vitro diagnostic test for the quantitative measurement of lactate dehydrogenase in human serum and plasma on the Dimension Vista System.Cat. No. K2054, Mat. No. 10464323

Reason

Siemens Healthcare Diagnostics has confirmed that Dimension Vista Lactate Dehydrogenase (LDI) may exhibit an erroneously elevated result when the preceding assay in the reaction cuvette is Lactic Acid (LA). The magnitude of the elevation may be variable. Both quality control and patient samples may be impacted. Incomplete removal of residual lactate dehydrogenase, a component of the LA reagent, has been identified as the cause of the potentially elevated result. A falsely elevated result for lactate dehydrogenase may lead to additional investigations to determine the etiology of tissue damage. Results would be correlated with clinical history and presentation in addition to other diagnostic laboratory testing.

Action

An Urgent Medical Device Correction were issued on October 11, 2018 to all Dimension Vista customers to notify them of the issue, the potential risk to health, and actions to be taken by the customer. Customers were instructed to: -Configure Dimension Vista System to eliminate the potential for cuvette carryover from lactic acid. -If unable to separate the assays due to lack of additional server based on the the Dimension Vista model, run Lactate Dehydrogenase method in duplicates and contact Siemens if either result is discordant. -Complete and return the Field Correction Effectiveness Check Form attached to the letter within 30 days. -If you have received any complaints of illness or adverse events associated with the product listed in Table 1 of the letter, immediately contact your local Siemens Customer Care Center or their local Siemens Technical Support Representative

Distribution

Worldwide-domestic nationwide and foreign distribution.

Quantity

13435