Equator Forced Air Convective Warming System, Reference Number EQ 5000

Recall: Z-0497-05 · Initiated December 15, 2004

Recall

Recall Number
Z-0497-05
Event Number
30703
Firm
Smiths Medical ASD, Inc.
FEI Number
1221261
Product Code
DWJ
Status
Terminated
Root Cause
Other
Initiated
December 15, 2004
Posted
February 3, 2005
Terminated
June 27, 2006
Address
160 Weymouth St, Rockland, MA, 02370-1136

Description

Equator Forced Air Convective Warming System, Reference Number EQ 5000

Reason

Warmer heater coil may come in contact with the filter housing, causing a small area of the plastic housing to melt

Action

Smiths Medical ASD notified customers by letter issued Fed''X on 12/15/04. Accounts are requested not to use the device and return the unit if the the unit has not yet been placed into service. If the unit has been placed into service and operated for more than one hour then the problem will not occur. EQ-5000 unit(s) will be replaced.

Distribution

Nationwide Foreign: Nederland BV, UK, Finland

Quantity

206 units