FDA Recall
Terminated
Equator Forced Air Convective Warming System, Reference Number EQ 5000
Recall: Z-0497-05
·
Initiated December 15, 2004
Recall
- Recall Number
- Z-0497-05
- Event Number
- 30703
- Firm
- Smiths Medical ASD, Inc.
- FEI Number
- 1221261
- Product Code
- DWJ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- December 15, 2004
- Posted
- February 3, 2005
- Terminated
- June 27, 2006
- Address
- 160 Weymouth St, Rockland, MA, 02370-1136
Description
Equator Forced Air Convective Warming System, Reference Number EQ 5000
Reason
Warmer heater coil may come in contact with the filter housing, causing a small area of the plastic housing to melt
Action
Smiths Medical ASD notified customers by letter issued Fed''X on 12/15/04. Accounts are requested not to use the device and return the unit if the the unit has not yet been placed into service. If the unit has been placed into service and operated for more than one hour then the problem will not occur. EQ-5000 unit(s) will be replaced.
Distribution
Nationwide Foreign: Nederland BV, UK, Finland
Quantity
206 units