FDA Recall Terminated

Hemosphere System with the following three components: HEM1 Advanced Monitor, HEMSGM10 (Swan-Ganz Module), and HEMOXSC100 (Oximetry Smart Cable). Product Usage: The HemoSphere Advanced Monitor (HEM1) is intended to be used in combination with a compatible Edwards hemodynamic monitoring technology module and/or cable and accompanying Edwards accessories and/or disposables. It provides Edwards hemodynamic monitoring technologies through modular connections. The HemoSphere Swan-Ganz Module (HEMSGM10) is intended to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting. It is a multi-use device that works in conjunction with the HemoSphere Advanced Monitor, patient CCO cable and compatible Swan-Ganz catheter to acquire and process temperature data supporting intermittent cardiac output (iCO), continuous cardiac output (CCO) and right ventricular end diastolic volume (EDV) monitoring. The HemoSphere Oximetry Smart Cable (HEMOXSC100) is intended to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting. It is a multi-use device that connects with a compatible Edwards monitor on one end and any approved Edwards oximetry catheter on the other end to continuously measure venous oxygen saturation by reflectance spectrophotometry. LEDs within the oximetry cable transmit light fiber optically to the distal end of the catheter. The amount of light absorbed, refracted, and reflected depends on the relative amounts of oxygenated and deoxygenated hemoglobin in the blood. This optical intensity data is gathered by the oximetry catheter, processed by the HemoSphere Oximetry Cable, and displayed on a compatible monitoring platform. Parameter output is mixed venous oxygen saturation (SvO2) or central venous oxygen saturation (ScvO2).

Recall: Z-0496-2018 · Initiated October 2, 2017

Recall

Recall Number
Z-0496-2018
Event Number
78442
Firm
Edwards Lifesciences, LLC
FEI Number
2015691
Product Code
DQK
Status
Terminated
Root Cause
Device Design
Initiated
October 2, 2017
Posted
January 16, 2018
Terminated
December 14, 2020
Address
1 Edwards Way, Irvine, CA, 92614-5688

Description

Hemosphere System with the following three components: HEM1 Advanced Monitor, HEMSGM10 (Swan-Ganz Module), and HEMOXSC100 (Oximetry Smart Cable). Product Usage: The HemoSphere Advanced Monitor (HEM1) is intended to be used in combination with a compatible Edwards hemodynamic monitoring technology module and/or cable and accompanying Edwards accessories and/or disposables. It provides Edwards hemodynamic monitoring technologies through modular connections. The HemoSphere Swan-Ganz Module (HEMSGM10) is intended to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting. It is a multi-use device that works in conjunction with the HemoSphere Advanced Monitor, patient CCO cable and compatible Swan-Ganz catheter to acquire and process temperature data supporting intermittent cardiac output (iCO), continuous cardiac output (CCO) and right ventricular end diastolic volume (EDV) monitoring. The HemoSphere Oximetry Smart Cable (HEMOXSC100) is intended to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting. It is a multi-use device that connects with a compatible Edwards monitor on one end and any approved Edwards oximetry catheter on the other end to continuously measure venous oxygen saturation by reflectance spectrophotometry. LEDs within the oximetry cable transmit light fiber optically to the distal end of the catheter. The amount of light absorbed, refracted, and reflected depends on the relative amounts of oxygenated and deoxygenated hemoglobin in the blood. This optical intensity data is gathered by the oximetry catheter, processed by the HemoSphere Oximetry Cable, and displayed on a compatible monitoring platform. Parameter output is mixed venous oxygen saturation (SvO2) or central venous oxygen saturation (ScvO2).

Reason

Pre-procedural issues related to software defects

Action

Consignees were sent recall notification letters on approximately 10/11/2017. The letter identified the affected product, problem and actions to be taken. instructions included coordination with the respective Edwards representative to schedule the software upgrade.

Distribution

US Nationwide Distributed in the states of AZ, CA, CO, GA, IL, MO, NM, OK, OR, SC, TX

Quantity

198 devices