FDA Recall Terminated

LigaSure Blunt Tip Laparoscopic Sealer/Divider, Nano-Coated, Product Number LF1837

Recall: Z-0495-2022 · Initiated December 6, 2021

Recall

Recall Number
Z-0495-2022
Event Number
89182
Firm
Covidien Llc
FEI Number
1282497
Product Code
GEI
Status
Terminated
Root Cause
Device Design
Initiated
December 6, 2021
Posted
January 13, 2022
Terminated
May 11, 2023
Address
15 Hampshire St, Mansfield, MA, 02048-1113

Description

LigaSure Blunt Tip Laparoscopic Sealer/Divider, Nano-Coated, Product Number LF1837

Reason

Customers reported that the device jaws were difficult to open or would not open following application on tissue. Use of a device with this potential assembly issue could result in the potential for bleeding, unintended tissue injury or loss, and delay in treatment while an alternate device is obtained.

Action

URGENT MEDICAL DEVICE RECALL notification letters dated 12/2021 were sent to customers. Customer Action Required: 1. Please immediately quarantine and discontinue use of the affected item code with the associated lot number listed above. 2. Please return affected product as indicated below. All products from the affected item code and associated lot number must be returned. 3. If you have distributed the product item code and lot number listed above, please promptly forward the information from this letter to those recipients. 4. Complete the Recalled Product Return Form even if you do not have inventory and email it to [email protected] We regret any inconvenience this may cause. We are committed to patient safety and appreciate your prompt attention to this matter. If you have any questions regarding this communication, please do not hesitate to contact Azi Hosseini on +61 428 464 838, or [email protected]

Distribution

Worldwide Distribution in the countries of Australia, Austria, Belgium, Czech Republic, France, Germany, Ireland, Kazakhstan, Netherlands, Spain, United Kingdom. No US distribution.

Quantity

504 devices