LIFEPAK 12 defibrillator/ monitors biphasic
Recall
- Recall Number
- Z-0495-05
- Event Number
- 30915
- Firm
- Medtronic Emergency Response Systems, Inc.
- FEI Number
- 3015876
- Product Code
- MKJ
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- January 15, 2005
- Posted
- February 3, 2005
- Terminated
- May 6, 2006
- Address
- 11811 Willows Rd Ne, Redmond, WA, 98052-2003
Description
LIFEPAK 12 defibrillator/ monitors biphasic
The LIFEPAK 12 biphasic defibrillator/ monitors may deliver inappropriate energy levels and fail to escalate energy levels when configured to do so.
The firm sent 4000 User Notification letters via FED-EX on 1/15/2005 and approximately 3000 on 1/17/2005. The User Notification letters informed consignees about the two complaints and described the firm''s findings from their investigation into the problem. The letters also requested the users check the default energy settings; report back to the firm any discrepancies found; and reset the discrepant settings which differed from the user programmed settings.
28,835 devices were shipped to 6874 domestic consignees.
28,835 devices distributed in the U.S.; 15,613 devices distributed internationally