FDA Recall Terminated

LIFEPAK 12 defibrillator/ monitors biphasic

Recall: Z-0495-05 · Initiated January 15, 2005

Recall

Recall Number
Z-0495-05
Event Number
30915
Firm
Medtronic Emergency Response Systems, Inc.
FEI Number
3015876
Product Code
MKJ
Status
Terminated
Root Cause
Software design
Initiated
January 15, 2005
Posted
February 3, 2005
Terminated
May 6, 2006
Address
11811 Willows Rd Ne, Redmond, WA, 98052-2003

Description

LIFEPAK 12 defibrillator/ monitors biphasic

Reason

The LIFEPAK 12 biphasic defibrillator/ monitors may deliver inappropriate energy levels and fail to escalate energy levels when configured to do so.

Action

The firm sent 4000 User Notification letters via FED-EX on 1/15/2005 and approximately 3000 on 1/17/2005. The User Notification letters informed consignees about the two complaints and described the firm''s findings from their investigation into the problem. The letters also requested the users check the default energy settings; report back to the firm any discrepancies found; and reset the discrepant settings which differed from the user programmed settings.

Distribution

28,835 devices were shipped to 6874 domestic consignees.

Quantity

28,835 devices distributed in the U.S.; 15,613 devices distributed internationally