FDA Recall Terminated

KIT URIN CUP 16X100 8.0 CONI UA YEL, Cat. No. 364989 BD Vacutainer Urine Cup Kit with UA Tube for Midstream Specimens

Recall: Z-0494-2019 · Initiated August 30, 2018

Recall

Recall Number
Z-0494-2019
Event Number
81148
Firm
Becton Dickinson & Company
FEI Number
2243072
Product Code
JTW
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
August 30, 2018
Terminated
October 21, 2020
Address
1 Becton Dr, Franklin Lakes, NJ, 07417-1815

Description

KIT URIN CUP 16X100 8.0 CONI UA YEL, Cat. No. 364989 BD Vacutainer Urine Cup Kit with UA Tube for Midstream Specimens

Reason

BD Vacutainer Urine Collection Cups may contribute to a slight elevation of Red Blood Cell (RBC) counts when used with the Sysmex UF Series Analyzers. BD recommends following Good Clinical Practice (GCP) by using microscopy to confirm an elevated RBC Count in a Urine sample as per CLSI GP16-3A. 29(4) and CAP Checklist on Urinalysis. Product will continue to ship to not impact critical patient care.

Action

On August 30, 2018, BD notified Sysmex via an emailed Urgent Medical Device Product Safety Alert letter. Sysmex was informed that BD had confirmed the product issue with BD Vacutainer Urine Collection Cups were used with Sysmex UF Series Analyzers. Sysmex was instructed to take the following actions: - Notify Sysmex customers of the potential for elevation of RBC counts - Complete the customer Resopnse form and email it to BD at [email protected] to confirm receipt of the Safety Alert. Product will continue to ship so as to not impact critical patient care. BD recommends following Good Clinical Practice (GCP) by performing microscopy to confirm an elevated RBC Count in a urine sample as per CLSI GP16-3A.29(4) and CAP Checklist on Urinalysis.

Distribution

Illinois

Quantity

Total: 29,982,150 BD cups (US); 97,800,900 BD cups (OUS)