FDA Recall Terminated

VERSAPOINT, Hysteroscopic system Connector Cable, Non-Sterile, Reusable (20 uses only). US Distribution/Manufactured for GYNECARE, A division of Ethicon, Inc., a Johnson & Johnson Company, Somerville, NJ 08876. The product subject of this recall is the connector cable which goes between the electrosurgical generator and the elctrode used for surgery. Specifically, the recall focuses on the hand grip on the electrode, formed by overmolding a polymer onto the cable, which deforms upon sterilization.

Recall: Z-0493-05 · Initiated December 16, 2004

Recall

Recall Number
Z-0493-05
Event Number
30711
Firm
Ethicon, Inc US
FEI Number
2210968
Product Code
HIH
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
December 16, 2004
Posted
February 2, 2005
Terminated
October 31, 2005
Address
Highway 22 West, Somerville, NJ, 08876

Description

VERSAPOINT, Hysteroscopic system Connector Cable, Non-Sterile, Reusable (20 uses only). US Distribution/Manufactured for GYNECARE, A division of Ethicon, Inc., a Johnson & Johnson Company, Somerville, NJ 08876. The product subject of this recall is the connector cable which goes between the electrosurgical generator and the elctrode used for surgery. Specifically, the recall focuses on the hand grip on the electrode, formed by overmolding a polymer onto the cable, which deforms upon sterilization.

Reason

Deformation of the Verspoint Hysteroscopic System Connector Cable hand grips when autoclaved causing possible non-sterile electrodes.

Action

The recall letter was mailed to the Director of Material Management and the Director of Operating Room on December 14, 2004.

Distribution

Ethicon distributed the devices to 166 hospitals in the US. There are no Government sales. Product sold to other countries was distributed directly from the manufacturerin England to Ethicon distributors outside the US. The Foreign countries with accounts are: Austria, Czech Republic, Greece, Italy, Benelux, Poland, Portugal, Slovenia, Spain, Sweden, France, Germany, Ireland, U.K. Canada, China, Mexico, Taiwan, Venezuela, Korea, Australia, Brazil, and India.

Quantity

891electrodes