TosohSorter 9tray, Product Code 022931and TosohSorter 19tray, Product Code 022932. Product Usage: The AIA-900 9tray Sorter and AIA-900 19tray Sorter supports the automation of the AIA-900 systems.
Recall
- Recall Number
- Z-0492-2018
- Event Number
- 78776
- Firm
- Tosoh Bioscience Inc
- FEI Number
- 3005529799
- Product Code
- KHO
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- June 11, 2017
- Terminated
- December 31, 2019
- Address
- 3600 Gantz Rd, Grove City, OH, 43123-1895
Description
TosohSorter 9tray, Product Code 022931and TosohSorter 19tray, Product Code 022932. Product Usage: The AIA-900 9tray Sorter and AIA-900 19tray Sorter supports the automation of the AIA-900 systems.
This recall has been initiated due to a potential malfunction of the photo interrupter (Pl). The potential malfunction has been attributed to aging and premature deterioration of the Pl collector current.
Customers were sent a recall notification by email on July 11, 2017 informing them that their local Field Service Engineer would contact them to make arrangements to proactively replace the affected parts. Follow up notifications will be by Federal Express with verification of delivery.
Worldwide Distribution - U.S. Nationwide - Latin America
565 total devices U.S.