FDA Recall Terminated

Terumo Advanced Perfusion System 1; Integrated (Sarns) Centrifugal System Control Unit; Catalog number 801046.

Recall: Z-0492-2007 · Initiated October 16, 2006

Recall

Recall Number
Z-0492-2007
Event Number
36496
Firm
Terumo Cardiovascular Systems Corp
FEI Number
1828100
Product Code
DTQ
Status
Terminated
Root Cause
Other
Initiated
October 16, 2006
Posted
February 14, 2007
Terminated
July 7, 2009
Address
6200 Jackson Rd, Ann Arbor, MI, 48103-9586

Description

Terumo Advanced Perfusion System 1; Integrated (Sarns) Centrifugal System Control Unit; Catalog number 801046.

Reason

The pump may lose communication with the central control monitor, causing the monitor to display a service message "System Computer Needs Service" and to become non-operational.

Action

Consignees were notified of the problem via letter dated 11/10/06 and informed that they would be contacted to schedule a service call to install software upgrades. Installation of software upgrades began at customer locations on 10/16/06.

Distribution

Worldwide distribution --- including USA and countries of Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.

Quantity

343