FDA Recall Terminated

AdvanDx E. coli/P. aeruginosa PNA FISH Culture Identification Kit for the diagnosis of E. coli and/or P. aeruginosa bacteremia Catalog Number: KT007

Recall: Z-0490-2011 · Initiated October 6, 2010

Recall

Recall Number
Z-0490-2011
Event Number
57002
Firm
AdvanDx, Inc.
FEI Number
3004080598
Product Code
JSS
Status
Terminated
Root Cause
Device Design
Initiated
October 6, 2010
Posted
November 29, 2010
Terminated
June 6, 2011
Address
10a Roessler Rd, Woburn, MA, 01801-6208

Description

AdvanDx E. coli/P. aeruginosa PNA FISH Culture Identification Kit for the diagnosis of E. coli and/or P. aeruginosa bacteremia Catalog Number: KT007

Reason

Strains of Acinetobacter radioresistens cross-react with both E. coli/P. aeruginosa PNA FISH and EK/P. aeruginosa PNA FISH to produce a false positive red signal.

Action

AdvanDx issued a customer letter dated October 6, 2010 with a PDF file of a new package insert and an acknowledgement form to fill out and return to AdvanDx. Customers were instructed to replace the package insert in any existing inventory with the enclosed package insert. Customers can contact AdvanDx technical support about this action at 1-866-376-0009.

Distribution

Worldwide Distribuition: Throughtout the US, and to the countries of Denmark, Italy, Sweden, and UK.

Quantity

260 units