AdvanDx E. coli/P. aeruginosa PNA FISH Culture Identification Kit for the diagnosis of E. coli and/or P. aeruginosa bacteremia Catalog Number: KT007
Recall
- Recall Number
- Z-0490-2011
- Event Number
- 57002
- Firm
- AdvanDx, Inc.
- FEI Number
- 3004080598
- Product Code
- JSS
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- October 6, 2010
- Posted
- November 29, 2010
- Terminated
- June 6, 2011
- Address
- 10a Roessler Rd, Woburn, MA, 01801-6208
Description
AdvanDx E. coli/P. aeruginosa PNA FISH Culture Identification Kit for the diagnosis of E. coli and/or P. aeruginosa bacteremia Catalog Number: KT007
Strains of Acinetobacter radioresistens cross-react with both E. coli/P. aeruginosa PNA FISH and EK/P. aeruginosa PNA FISH to produce a false positive red signal.
AdvanDx issued a customer letter dated October 6, 2010 with a PDF file of a new package insert and an acknowledgement form to fill out and return to AdvanDx. Customers were instructed to replace the package insert in any existing inventory with the enclosed package insert. Customers can contact AdvanDx technical support about this action at 1-866-376-0009.
Worldwide Distribuition: Throughtout the US, and to the countries of Denmark, Italy, Sweden, and UK.
260 units