FDA Recall Terminated

Minicon Clinical Sample Concentrators for In Vitro Diagnostic Use 9031 (Model: Minicon B-15) & 9051 (Model: Minicon CS-15) Minicon concentrators are non-sterile, disposable, multiwall ultrafiltration devices. Minicon clinical sample concentrators are for in vitro diagnostic use and are intended for concentrating serum, urine, cerebrospinal fluid, and other body fluids prior to analysis.

Recall: Z-0489-2019 · Initiated May 14, 2018

Recall

Recall Number
Z-0489-2019
Event Number
81122
Firm
Merck Millipore Ltd. Tullagreen Carrigtwohill County Cork Ireland
FEI Number
1000532580
Product Code
JJH
Status
Terminated
Root Cause
Process control
Initiated
May 14, 2018
Terminated
August 5, 2021

Description

Minicon Clinical Sample Concentrators for In Vitro Diagnostic Use 9031 (Model: Minicon B-15) & 9051 (Model: Minicon CS-15) Minicon concentrators are non-sterile, disposable, multiwall ultrafiltration devices. Minicon clinical sample concentrators are for in vitro diagnostic use and are intended for concentrating serum, urine, cerebrospinal fluid, and other body fluids prior to analysis.

Reason

In vitro diagnostic devices, used greater than 6 months after date of manufacture, may have increased concentration times of greater than the specification of 3 hours and up to 24 hours, which may lead to delays in carrying out the subsequent analysis tests.

Action

On May 14, 2018, the firm sent an "Urgent Field Safety Notice" to customers via email and following up with a hard-copy letter on June 6, 2018 via United States Postal Services. The letter described the product, problem and actions to be taken. The corrective action being taken by the firm is to reduce the shelf life of the device from 12 months to 6 months. The customers are being instructed to immediately examine their inventory to determine if they have any remaining impacted product and forward a copy of this letter to the personnel in your organization that utilize this product and/or need to be made aware of this issue. In addition, if the customer has further distributed this product, please identify its customers and notify them at once of this issue. The notification to their customers should include a copy of this notification and attachment 2 Customers are instructed to please stop use of these lots if older than 6 months and destroy any remaining inventory older than 6 months, and to complete and return the receipt confirmation form to [email protected]. If you would like more information regarding this notification, please contact us at [email protected] or go to our website to find the phone number of your local office at www.emdmillipore.com/offices.

Distribution

Worldwide Distribution: US (nationwide) to states of: AZ, CA, CT, DC, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NJ, NY, OH, OK,OR, PA, RI, TN, TX, VA, VT, WA and WI; and countries of: Canada and Australia.

Quantity

1360 units