FDA Recall Open, Classified

Ascom Telligence Nurse Call System

Recall: Z-0486-2019 · Initiated November 2, 2018

Recall

Recall Number
Z-0486-2019
Event Number
81546
Firm
Ascom (US) Inc.
FEI Number
3005334549
Product Code
IQA
Status
Open, Classified
Root Cause
Software Manufacturing/Software Deployment
Initiated
November 2, 2018
Address
9024 Town Center Parkway, Lakewood Ranch, FL, 34202-4159

Description

Ascom Telligence Nurse Call System

Reason

There is an issue that could cause the Telligence System Gateway II, NGGTWY2 H, to stop communicating to the hallway devices connected to it. When this occurs, the dome lights connected to the Gateway II Ports A and B will start flashing, indicating that they are no longer communicating to the Gateway II. If this occurs, device failures that WOULD otherwise be annunciated WILL NOT be indicated at staff consoles that are configured to display failures.

Action

Ascom Partners / Distributors were contacted by e-Mail, USPS Mail and Phone as necessary on 11/02/2018 to determine where the Telligence 5.0 Product is installed.

Distribution

AL, CA, CO, GA, IN, KY, LA, ME, MI, MO, MN, MS, NE, NV, OH, PA, RI, TX UT, VA, WI

Quantity

57 units