Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); CFN: SIGSDL45CTVT;
Recall
- Recall Number
- Z-0485-2026
- Event Number
- 97792
- FEI Number
- 1219930
- Product Code
- GDW
- Status
- Open, Classified
- Root Cause
- Process control
- Initiated
- September 25, 2025
- Posted
- November 7, 2025
- Address
- LP 60 Middletown Ave, Covidien, North Haven, CT, 06473-3908
Description
Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); CFN: SIGSDL45CTVT;
A manufacturing issue could lead to a malformed cover tube press which can lead to internal components' movement during firing - leading to the potential for uncontrolled motion.
On September 25, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Customer actions: Identify and quarantine all unused impacted product listed. Return all quarantined devices to Medtronic. If purchased from a distributor, contact your distributor directly to arrange for the return of the devices back to your distributor. Please complete and return the enclosed Customer Confirmation Form. Pass on this notice to all those who need to be aware within your organization or to any organization where the listed affected lots may have been transferred or distributed.
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Chile, Estonia, Finland, France, Germany, Greece, Hungary, Israel, Italy, Japan, Kuwait, Malaysia, Netherlands, Poland, Portugal, Romania, Saudi Arabia, Slovakia, Spain, United Kingdom.
3263 units