Fisher & Paykel IW930 CosyCot Infant Radiant Warmer (with integrated bed) fitted with 9001W005J Elevator Module
Recall
- Recall Number
- Z-0485-2007
- Event Number
- 37003
- Firm
- Fisher & Paykel Healthcare Ltd 15 Maurice Paykel Place Auckland New Zealand
- FEI Number
- 3002806988
- Product Code
- FMT
- Status
- Terminated
- Root Cause
- Other
- Initiated
- October 9, 2006
- Posted
- January 11, 2007
- Terminated
- April 26, 2012
Description
Fisher & Paykel IW930 CosyCot Infant Radiant Warmer (with integrated bed) fitted with 9001W005J Elevator Module
This firm has received Product Complaint Reports concerning bending of the pivot plate. The elevator base of the CosyCot Infant Warmer incorporates a pivot plate which forms part of the mechanism allowing the CosyCot to raise and lower in height. A bent pivot plate generally results in the Warmer being unable to elevate properly. Other component failures may result.
Fisher & Paykel Healthcare commenced sending out Product Correction Notifications letter via U.S. Postal Service certified mail October 9, 2006. This Product Correction Notice explained the nature of the problem and include a reminder of the maximum recommended weight of accessories. Two FAX Response Forms were provided, Part A for initial receipt of the Product Correction Notice and Part B for confirmation of the completed upgrade.
Nationwide and Canada
273 units in US & 9 in Canada