FDA Recall Terminated

Fisher & Paykel IW930 CosyCot Infant Radiant Warmer (with integrated bed) fitted with 9001W005J Elevator Module

Recall: Z-0485-2007 · Initiated October 9, 2006

Recall

Recall Number
Z-0485-2007
Event Number
37003
Firm
Fisher & Paykel Healthcare Ltd 15 Maurice Paykel Place Auckland New Zealand
FEI Number
3002806988
Product Code
FMT
Status
Terminated
Root Cause
Other
Initiated
October 9, 2006
Posted
January 11, 2007
Terminated
April 26, 2012

Description

Fisher & Paykel IW930 CosyCot Infant Radiant Warmer (with integrated bed) fitted with 9001W005J Elevator Module

Reason

This firm has received Product Complaint Reports concerning bending of the pivot plate. The elevator base of the CosyCot Infant Warmer incorporates a pivot plate which forms part of the mechanism allowing the CosyCot to raise and lower in height. A bent pivot plate generally results in the Warmer being unable to elevate properly. Other component failures may result.

Action

Fisher & Paykel Healthcare commenced sending out Product Correction Notifications letter via U.S. Postal Service certified mail October 9, 2006. This Product Correction Notice explained the nature of the problem and include a reminder of the maximum recommended weight of accessories. Two FAX Response Forms were provided, Part A for initial receipt of the Product Correction Notice and Part B for confirmation of the completed upgrade.

Distribution

Nationwide and Canada

Quantity

273 units in US & 9 in Canada