FDA Recall Terminated

Panda Wall Mount warmer. Infant radiant warmer

Recall: Z-0484-2019 · Initiated September 25, 2018

Recall

Recall Number
Z-0484-2019
Event Number
81224
Firm
GE Healthcare, LLC
FEI Number
2126677
Product Code
FMT
Status
Terminated
Root Cause
Process design
Initiated
September 25, 2018
Terminated
July 13, 2020
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

Panda Wall Mount warmer. Infant radiant warmer

Reason

Potential electric leakage affecting the user.

Action

GE Healthcare notified customers on about 09/25/2018 via URGENT MEDICAL DEVICE CORRECTION letter. The letter indicated that the device could continue to be used along with good clinical practices. GE Healthcare will visit each location to perform the necessary testing. Until GE Healthcare can visit each site, instructions are to disconnect devices that use the USB (Giraffe OmniBed Carestation and Giraffe Incubator Carestation only) and the RS232 Nurse call ports, and inspect all power cords to ensure there is no damage to the cord or ground pin. GE Healtchare will perform necessary electrical safety test on all affected products and correct if required. A GE Healthcare representative will contact each customer to arrange for this test. For questions or concerns, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Distribution

Worldwide distribution. US Nationwide, Algeria, Australia, Austria, Belgium, Canada, Chile, China, Costa Rica, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Honduras, India, Indonesia, Ireland, Italy, Japan, Kazakhstan, Korea, Lithuania, Mexico, Netherlands, New Zealand, Norway, Pakistan, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam.

Quantity

1,724 devices total