HeartStart FR2+ AED with ECG Display, Model M3860, Philips brand, configurable manual charge in advance mode. The label of the device states: "HEARTSTART...FR2+ Automated External Defibrillator...Manufactured by: Philips Medical Systems Seattle, WA 98121".
Recall
- Recall Number
- Z-0483-2011
- Event Number
- 56816
- Firm
- Philips Medical Systems
- FEI Number
- 1000524572
- Product Code
- MKJ
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- September 22, 2010
- Posted
- November 29, 2010
- Terminated
- September 5, 2013
- Address
- 2301 5th Ave, Ste 200, Seattle, WA, 98121
Description
HeartStart FR2+ AED with ECG Display, Model M3860, Philips brand, configurable manual charge in advance mode. The label of the device states: "HEARTSTART...FR2+ Automated External Defibrillator...Manufactured by: Philips Medical Systems Seattle, WA 98121".
A flash memory component in the HeartStart FR2+ and Automated External Defibrillator was improperly relabeled by a third party. Therefore the component could not be identified by its source, quality, and date of manufacture.
On 10/5/10, Philips began sending out the Voluntary Medical Device Removal notification letter to their consignees. This letter informs the consignees that Philips Healthcare is conducting a voluntary product removal of a limited number of HeartStartFR2+ automated external defibrillators (AEDs), model numbers M3860A, M3861A, M3849A, and M3841A, manufactured by Philips and shipped between March 2007 and August 2010. The owners of affected AEDs are advised to run a battery insertion test. If it passes, the AED should remain in service until its replacement is received. Philips will be replacing affected units. The consignees can visit the firm's website www.philips.com/FR2AEDAction or contact Philips at 1-800-263-3342 for any information or support concerning the issue.
Devices were distributed to hospitals, medical facilities, and distributors through out the US and to foreign consignees. Foreign consignees including countries: Algeria, Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Ecuador, El Salvador, Finland, France, French Guiana, Georgia, Germany, Ghana, Gibraltar, Guatemala, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Japan, Jordan, Kampuchea, Kenya, Korea, Latvia, Lebanon, Libya, Luxembourg, Malaysia, Malta & Gozo, Mexico, Netherland, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestinian Territory, Panama, Peru, Philippines, Portugal, Puerto Rico, Qatar, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, United Arab Emirates, United Kingdom, Vietnam, and Zimbabwe.
10359 units