FDA Recall Terminated

Animas ezManager Max Diabetes Management System version 2.0.9

Recall: Z-0481-2010 · Initiated November 2, 2009

Recall

Recall Number
Z-0481-2010
Event Number
53748
Firm
Animas Corporation
FEI Number
3002920072
Product Code
MRZ
Status
Terminated
Root Cause
Software design
Initiated
November 2, 2009
Posted
December 3, 2009
Terminated
October 27, 2010
Address
200 Lawrence Dr, West Chester, PA, 19380-3428

Description

Animas ezManager Max Diabetes Management System version 2.0.9

Reason

Printed hard copy of results is not the same as the system screens- carbohydrate to insulin ratio incorrect.

Action

Animas issued an Urgent Medical Device Correction letter dated 11/2/09 to the pump user and another letter to the health care professional that prescribed the pump. The letters explain the issue. The patient letter is accompanied with a CD that contains the upgraded software and instructions to follow.

Distribution

Nationwide distribution, and Canada, Cayman Islands, and Denmark.

Quantity

20,648 units