FDA Recall Terminated

Maquet Accessories, Connection Bracket. Catalogue # 1180.36A0. An accessory component for the MAGNUS operating table system. Used to mount head positioning accessories.

Recall: Z-0480-2010 · Initiated October 22, 2009

Recall

Recall Number
Z-0480-2010
Event Number
53612
Firm
Maquet Inc.
FEI Number
3008355164
Product Code
FWZ
Status
Terminated
Root Cause
Device Design
Initiated
October 22, 2009
Posted
December 2, 2009
Terminated
May 18, 2010
Address
45 Barbour Pond Drive, Wayne, NJ, 07470

Description

Maquet Accessories, Connection Bracket. Catalogue # 1180.36A0. An accessory component for the MAGNUS operating table system. Used to mount head positioning accessories.

Reason

Product monitoring has determined that the connection bracket shows potential mechanical play at the OR table interface for some procedures. This is due to the hole in the lock washer, which is responsible for locking the accessory to the OR table, being too large.

Action

Urgent Device Correction Letters were sent by Federal Express to affected US hospitals on October 22, 2009. The letter stated the technical description, potential hazard, and the recommended precaution. The letter asked customers to immediately verify whether their Connection Bracket is affected thru the FaxBack provided. Questions or need for additional information can be directed to customers' local MAQUET representative.

Distribution

Nationwide Distribution -- USA, including states of SC and PA.

Quantity

9 units in US.