BHM Kwiktrak Fixed Gate component of the Kwiktrak rail system used with the Maxi Sky 600 Ceiling Patient Lift; Manufactured by BMH Medical Inc., 2001 Tanguay Street , Magog, Quebec, Canada J1X 5Y5. The fixed gate, part #700.11550, was sold in two kits: a) Part #700.11500 Kwiktrak Gate Kit composed of the fixed gate and a 3-contact interface kit part #700.15515, and b) Part #700-11505 Double Gate Kit composed of two fixed gates and 3-contact interface kits

Recall: Z-0479-2007 · Initiated October 23, 2006

Recall

Recall Number
Z-0479-2007
Event Number
36636
Firm
Arjo, Inc.
FEI Number
1419652
Product Code
FSA
Status
Terminated
Root Cause
Other
Initiated
October 23, 2006
Posted
January 11, 2007
Terminated
January 9, 2010
Address
50 Gary Ave, Ste A, Roselle, IL, 60172-1684

Description

BHM Kwiktrak Fixed Gate component of the Kwiktrak rail system used with the Maxi Sky 600 Ceiling Patient Lift; Manufactured by BMH Medical Inc., 2001 Tanguay Street , Magog, Quebec, Canada J1X 5Y5. The fixed gate, part #700.11550, was sold in two kits: a) Part #700.11500 Kwiktrak Gate Kit composed of the fixed gate and a 3-contact interface kit part #700.15515, and b) Part #700-11505 Double Gate Kit composed of two fixed gates and 3-contact interface kits

Reason

The Kwiktrak Gate stopper may fail and allow the ceiling lift to pass through to the end of the track and fall to the floor.

Action

Arjo Inc. USA sent recall letters to the direct accounts on 10/30/06, informing them that if the ceiling lift is moved too fast or too forcefully into a closed gate, the gate stopper can be damaged, causing its function to fail. One consequence can be that the gate may remain open, allowing the ceiling lift to pass through to the end of the track. The letters requested the customer to visually inspect their Kwiktrak Gate installation used with their Maxisky-600 Ceiling Lift to determine if there are any signs of impact or burr marks on the closed gate stopper. Illustrations of a damaged and normal gate stopper were included. The customers were requested to stop use of the Kwiktrak track system if any signs of impact or burr marks are found, and contact the Arjo Quality Department at 1-800-323-1245, ext. 6140 for special instructions. If no signs of damage were found, the customers were reminded to make sure that they manually drive the ceiling lift in a gentler manner to prevent damage to the gate stopper, making sure that all personnel using the ceiling lifts are properly trained. The accounts were informed that Arjo personnel will contact them in the next few weeks to install an improved system that will eliminate the risks of failure.

Distribution

Michigan, Ohio and Illinois

Quantity

3 gate systems