FDA Recall Terminated

Insulia Diabetes Management Companion

Recall: Z-0478-2019 · Initiated October 8, 2018

Recall

Recall Number
Z-0478-2019
Event Number
81508
Firm
Voluntis 22 Quai Gallieni Suresnes France
FEI Number
3013530341
Product Code
NDC
Status
Terminated
Root Cause
Software in the Use Environment
Initiated
October 8, 2018
Terminated
March 12, 2019

Description

Insulia Diabetes Management Companion

Reason

The firm identified an issue with the basal calculator identified on the Android version of the Insulia application. The bug could lead from low-impact to high-impact hypoglycemia depending on the circumstances. A new update of Insulia is now available (version 1.6.42), which resolves this issue.

Action

On October 15, 2018, patients were notified that they could download the new version of the application from the Google Play Store. The notification was made through an automated message displayed on the application screen of all impacted patients requesting them to update their Insulia application. Patients were informed that the new version of the app resolved a an issue with the basal calculator identified on the Android version of the application. Patients who do not upgrade to the new version by November 9, 2018 will be blocked from using the app until they upgrade. For technical assistance or if you have any questions, please contact the Insulia support hotline.

Distribution

Distributed to accounts in Arizona, Masachusetts, Michigan, Ohio, Tennessee, and Vermont. International distribution to Canada and France (however, French customers are not affected by this recall).

Quantity

14