FDA Recall Terminated

AMSCO Evolution steam sterilizer Transfer Carriage. Product Usage: Mobile, wheeled transportation unit utilized to carry, load, & unload the Loading Car into and out of steam sterilizer units.

Recall: Z-0478-2014 · Initiated October 18, 2013

Recall

Recall Number
Z-0478-2014
Event Number
66874
Firm
Steris Corporation
FEI Number
1527821
Product Code
FLE
Status
Terminated
Root Cause
Device Design
Initiated
October 18, 2013
Posted
December 11, 2013
Terminated
May 13, 2014
Address
5960 Heisley Rd, Mentor, OH, 44060-1834

Description

AMSCO Evolution steam sterilizer Transfer Carriage. Product Usage: Mobile, wheeled transportation unit utilized to carry, load, & unload the Loading Car into and out of steam sterilizer units.

Reason

The firm discovered that unlatching can occur when overloaded transfer carriages are pulled or maneuvered incorrectly or with excessive force.

Action

Steris sent an Urgent Voluntary Field Correction Notice letter dated October 18, 2013 to their customers. The letter identified the affected product, problem, and actions to be taken. The letter informed customers that a Steris Service Representative will contact affected facility to arrange for Steris to upgrade the Transfer Carriage(s). For questions contact Steris Field Service Dispatch at 1-800-333-8828.

Distribution

Worldwide Distribution - USA Nationwide in the states of: AL, AZ, AR, CA, CO, CT, DE, FL, GA, ID, IL, IN, KS, KY, LA, ME, MD, MA, MI, MN, MI, MO, MT, NE, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY and the District of Columbia including the countries of: Brazil, Canada, Chile, Guam, Mexico, Saudi Arabia, United Arab Emirates & Venezuela.

Quantity

936 units