Minstrel Patient Lift with SR weighing scale; a non-AC-powered patient lift; Arjo, Inc., 50 N. Gary Ave., Roselle, IL 60172; model number HMB002-US

Recall: Z-0478-2007 · Initiated August 31, 2006

Recall

Recall Number
Z-0478-2007
Event Number
36187
Firm
Arjo, Inc.
FEI Number
1419652
Product Code
FNG
Status
Terminated
Root Cause
Other
Initiated
August 31, 2006
Posted
January 11, 2007
Terminated
October 16, 2008
Address
50 Gary Ave, Roselle, IL, 60172-1605

Description

Minstrel Patient Lift with SR weighing scale; a non-AC-powered patient lift; Arjo, Inc., 50 N. Gary Ave., Roselle, IL 60172; model number HMB002-US

Reason

The hanger bar assembly may detach from the lift during use due to fatigue failure of the bolt.

Action

Arjo sent an Urgent Device Field Correction Recall letter dated 8/31/06 to all affected end users to the attention of the Administrator/Risk Manager via FedEx on the same date. The letters informed the accounts of the potential for the hanger bar to detach from the scale due to fatigue failure of the pivot bolt. The accounts were advised that an Arjo Field Service Technician would contact them within five business days to schedule a visit to the account to perform an upgrade of their Minstrel lifts. Any questions were directed to Arjo Quality Department at 800-323-1245, ext. 6140.

Distribution

Nationwide

Quantity

76 lifts