FDA Recall
Terminated
Versys Hip System femoral stem, fiber metal midcoat, collared, 12/14 neck taper, low head center, size 15, standard body. Cat. No. 7841-15-10.
Recall: Z-0476-03
·
Initiated January 7, 2003
Recall
- Recall Number
- Z-0476-03
- Event Number
- 25375
- Firm
- Zimmer Inc.
- FEI Number
- 1000220733
- Product Code
- LPH
- Status
- Terminated
- Root Cause
- Other
- Initiated
- January 7, 2003
- Posted
- January 28, 2003
- Terminated
- February 9, 2004
- Address
- 345 E. Main St., Warsaw, IN, 46580
Description
Versys Hip System femoral stem, fiber metal midcoat, collared, 12/14 neck taper, low head center, size 15, standard body. Cat. No. 7841-15-10.
Reason
The femoral heads will not seat onto the taper of the hip stem, as the head impinges on the radius undercut below the taper.
Action
Recall letters dated 1/7/03 were sent to each U.S. consignee via certified mail, requesting that they contact Zimmer for pick-up of the products if any was in stock. Ex-U.S. consignees were notified via e-mail.
Distribution
United States, Australia and Canada.