FDA Recall Terminated

Total Toe Instrument Kit, Labeled in part: "Instrument Set, 15mm Toe, Dorsal Flange Combi" Part Number: 9000-3002 The component in question is the hex driver (part number 2009-4001).

Recall: Z-0475-2016 · Initiated June 9, 2014

Recall

Recall Number
Z-0475-2016
Event Number
72695
Firm
Arthrosurface, Inc.
FEI Number
3004154314
Product Code
LZJ
Status
Terminated
Root Cause
Process change control
Initiated
June 9, 2014
Posted
December 23, 2015
Terminated
February 5, 2016
Address
28 Forge Pkwy, Franklin, MA, 02038

Description

Total Toe Instrument Kit, Labeled in part: "Instrument Set, 15mm Toe, Dorsal Flange Combi" Part Number: 9000-3002 The component in question is the hex driver (part number 2009-4001).

Reason

An oversight in the implementation of the design change record resulted in outdated components being shipped to customers. The component did not pass a half-cycle sterilization.

Action

The distributors were verbally notified, and replacement Rev F Hex Drivers sent to the locations of the kits. The kits had been moved in preparation for upcoming surgeries, but the Drivers were replaced prior to any additional surgery taking place. No affected product remains in the field. For further questions, please call (508) 520-3003.

Distribution

US Distribution to the states of : PA, WV and CA.

Quantity

3 units