FDA Recall Terminated

Versys Hip System femoral stem, fiber metal midcoat, collared, 12/14 neck taper, low head center, size 12, standard body. Cat. No. 7841-12-10.

Recall: Z-0473-03 · Initiated January 7, 2003

Recall

Recall Number
Z-0473-03
Event Number
25375
Firm
Zimmer Inc.
FEI Number
1000220733
Product Code
LPH
Status
Terminated
Root Cause
Other
Initiated
January 7, 2003
Posted
January 28, 2003
Terminated
February 9, 2004
Address
345 E. Main St., Warsaw, IN, 46580

Description

Versys Hip System femoral stem, fiber metal midcoat, collared, 12/14 neck taper, low head center, size 12, standard body. Cat. No. 7841-12-10.

Reason

The femoral heads will not seat onto the taper of the hip stem, as the head impinges on the radius undercut below the taper.

Action

Recall letters dated 1/7/03 were sent to each U.S. consignee via certified mail, requesting that they contact Zimmer for pick-up of the products if any was in stock. Ex-U.S. consignees were notified via e-mail.

Distribution

United States, Australia and Canada.