Green Mamba Suture Passer

Recall: Z-0472-2016 · Initiated December 2, 2015

Recall

Recall Number
Z-0472-2016
Event Number
72829
Firm
Biomet, Inc.
FEI Number
1825034
Product Code
HWQ
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
December 2, 2015
Terminated
October 27, 2016
Address
56 E Bell Dr, Warsaw, IN, 46582-6989

Description

Green Mamba Suture Passer

Reason

There may be a separation and an insufficient weld between the spring and washer. The weld between the spring and washer may fracture, causing the Mamba Suture Passer instruments jaw to not actuate properly. 6 complaints have been received in which 3 have been reported with a delay up to 10 minutes.

Action

On 12/2/2015, URGENT MEDICAL DEVICE RECALL NOTICE notifications were sent to the affected distributors and hospital risk managers via courier with instructions to locate and return the affected products. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Questions related to this notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.

Distribution

FL, NJ, WI, UT, TX, KY, KS, OH, GA, LA, AL, NY, NC, AR, MO, VA, CA, MN, IN, HI

Quantity

13