FDA Recall Terminated

BrightMatter Guide, Product Number SYN-0026, Versions: 1.5.0, 1.5.1 and 1.5.2 Product Usage: BrightMatter Guide can navigate a third-party port (NICO BrainPath device) by attaching Synaptive s custom Tracking Array to the port directly. As the BrainPath is navigable in two configurations (sheath only / sheath and obturator), the tip of the obturator is represented in the software with a tip extension beyond the bottom of the sheath. Note sheath and cannula are used interchangeably in this document.

Recall: Z-0470-2019 · Initiated October 9, 2018

Recall

Recall Number
Z-0470-2019
Event Number
81317
Firm
Synaptive Medical Inc
FEI Number
3012075008
Product Code
HAW
Status
Terminated
Root Cause
Software design
Initiated
October 9, 2018
Posted
November 26, 2018
Terminated
August 12, 2021
Address
555 Richmond St West, Toronto Canada Suite 800

Description

BrightMatter Guide, Product Number SYN-0026, Versions: 1.5.0, 1.5.1 and 1.5.2 Product Usage: BrightMatter Guide can navigate a third-party port (NICO BrainPath device) by attaching Synaptive s custom Tracking Array to the port directly. As the BrainPath is navigable in two configurations (sheath only / sheath and obturator), the tip of the obturator is represented in the software with a tip extension beyond the bottom of the sheath. Note sheath and cannula are used interchangeably in this document.

Reason

This recall has been initiated due to a software defect found in the Guide System software when used with NICO BrainPath ports under certain circumstances. When one trajectory is set with a blue port and another trajectory is set with a gold port, the software defect is triggered when the user switches between these trajectories during a surgical procedure.

Action

Synaptive Medical sent an Urgent Medical Device Recall letter dated October 9, 2018 via registered priority mail. The letter identified the affected product, problem and actions to be taken until a software patch is released. Customers were requested to complete and return the enclosed Acknowledgement and Receipt form to Synaptive Medical Inc. For questions call 1-844-462-7246.

Distribution

Worldwide - US Nationwide Distribution and the country of Canada

Quantity

37 units