FDA Recall Open, Classified

Radiography 7300 C; Model Number: 712037;

Recall: Z-0466-2026 · Initiated October 1, 2025

Recall

Recall Number
Z-0466-2026
Event Number
97748
Firm
Philips Medical Systems DMC GmbH Development And Manufacturing Ctr. Rontgenstr. 24 Hamburg Germany
FEI Number
3003768251
Product Code
KPR
Status
Open, Classified
Root Cause
Device Design
Initiated
October 1, 2025
Posted
November 7, 2025

Description

Radiography 7300 C; Model Number: 712037;

Reason

Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/physical exertion required for manual movement) due to potential manufacturing issues.

Action

On October 1, 2025, "URGENT MEDICAL DEVICE CORRECTION" letters dated September 30, 2025, were sent to customers. The letters described the problem, product and actions to be taken. Actions to be taken: It is safe to continue to use the system(s) in accordance with the intended use. It is recommended that customers use the motorized movement option until the technical solution has been implemented. The system may experience motor drive errors if the force for movement exceeds 110N, which restricts the motorized movement. If the motorized movement does not function, customers should contact their local service representative. B. Circulate this notice to all users of this device so that they are aware of the issue and associated hazard/harm. C. Please retain this Urgent Medical Device Correction letter with your system(s) until the solution is installed; ensure the notice is in a place likely to be seen/viewed. 5. Actions planned by Philips to correct the problem. Philips will provide the technical solution details when finalized. This notice has been reported to the appropriate Regulatory Agencies. If you need any further information or support concerning this issue, please contact your local Philips representative. For North America, contact the Customer Care Solutions Center (1-800-722-9377, 8AM 8PM EST, Monday-Friday).

Distribution

Worldwide Distribution: US (Nationwide) Distribution and OUS (International) countries of: Australia, Austria, Bahrain, Belgium, Bosnia-Herz., Canada, Chile, Croatia, Denmark, Estonia, Finland, France, Germany, Ghana, Gibraltar, Guadeloupe, Indonesia, Iraq, Israel, Italy, Jordan, Kuwait, Kyrgyzstan, Latvia, Lithuania, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, R¿union, Samoa, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, T¿rkiye, Turkmenistan, United Kingdom, Utd.Arab.Emir.,

Quantity

284 units